Select Page

Comparative Physiology @ Guelph

Cleaning DEAE Sepharose Column

After Every Cycle

  • Run 2 column volumes of 1M NaCl Elution Buffer through
  • Run 3 column volumes of Running Buffer through to re-equilibrate

After Every 10 Cycles

  • Run one half column volume of 2M NaCl through in 10-15 minutes
  • Run one column volume of water
  • Run 1M NaOH for 1-2 hours
  • Run one column volume of water
  • Run 4 bed volumes of 70% Ethanol or 3% Isopropanol (1 hour)
  • Run one column volume water
  • Run 3 column volumes of Running Buffer to re-equilibrate